Please feel free to contact us in English at the following hot line. 0086+20+84528103

Temperature Chamber

Temperature Chamber-GPH-20

Temperature Chamber-GPH-H30

Temperature Chamber-GPH series(Horizontal type)

Our temperature Chambers are suitable for a broad range of applications from high-temperature testing, to heat treatment or drying for production lines,which widely used in aviation, automotive, electronic parts, materials, energy, chemical, medical, pharmaceutical and other industries.

Contact

Features:

1.High performance chamber

A temperature-indication controller with an advanced PID operation, and a special chamber configuration provide unmatched oven performance. Temperature uniformity, temperature constancy, temperature heat-up rate, and temperature recovery time are performed with the upmost reliability.

2.Several volumes are available in horizontal models.

3.Standard program instrumentation can suit various applications:

·Features programmed operation with operational settings such as constant mode and automatic start/stop. Suitable for heat treatment, drying, and similar production-line applications.

·Suitable for a range of applications from temperature-characteristics testing to heat treatment and drying. Programmed operation now allows storing 10 patterns, each up to 20 steps. Provides a wide range of functions, including temperature ramp settings and a maximum of 999 repeat cycles.

4.Easy setup with on-screen display.

5.Interface (option)for device communication can be selected between RS-485 and RS-232C.

6.Energy Saver Duct (PH series)

Energy saving approximately 20% by heat recycling from exhaust through the duct to maintain temperature while damper opens (PH series).

7.Safety measure

Triple safety mechanisms are employed for excessive overheating.

8.Core parts are inported from Japan and America,which are same with the ESPEC CORP.

SPECIFICATIONS

Description Model Temp. Range Hum. Control Hum. range rh Volume Inside dimensions in mm (WxHxD) Outside dimensions in mm (excl. projection)
Temperature Chamber(horizontal) GPH-10 (RT+20℃)~+200℃ NO 91L 450×450×450 1040×820×635
Temperature Chamber(horizontal) GPH-20 (RT+20℃)~+200℃ NO 216L 600×600×600 1190×970×785
Temperature Chamber(horizontal) GPH-30 (RT+20℃)~+200℃ NO 512L 800×800×800 1500×1210×1065
Temperature Chamber(horizontal) GPH-40 (RT+20℃)~+200℃ NO 1000L 1000×1000×1000 1730×1480×1275
Temperature Chamber(horizontal) GPH-H10 (RT+20℃)~+300℃ NO 91L 450×450×450 1040×820×635
Temperature Chamber(horizontal) GPH-H20 (RT+20℃)~+300℃ NO 216L 600×600×600 1190×970×785
Temperature Chamber(horizontal) GPH-H30 (RT+20℃)~+300℃ NO 512L 800×800×800 1500×1210×1065
Temperature Chamber(horizontal) GPH-H40 (RT+20℃)~+300℃ NO 1000L 1000×1000×1000 1730×1480×1275

※The temperature chamber's performance values are based on GB/T5170.2,and IEC60068-3-5.And humidity chamber's performance values are based on GB/T5170.5 and IEC60068-3-6.Performance figures aregiven for a +23℃ ambient temperature, relative humidity 65±20% RH, rated voltage, with no specimens inside the test area.

  • Exhaust port flange

    Flange connects an exhaust duct to the chamber to exhaust hot from the chamber.
    External diameter 87 mm.

  • Casters

    Installed for mobility.
    Specifications:Adjustable type 4pcs/set.(Including leveling feet 4pcs).

  • Recorder output terminal (Temp)

    Outputs chamber temperature .

  • Temp recorder wiring

    Preparation of a power cable,temperature sensor and a grounding wire for additional installation in the future.

  • Potable temp recorder (Paperless)

    Records temperature and humidity of each section such as the temperature inside the chamber.

  • Communication Function

    Communication ports to connect the chamber to a PC.
    Alternatives:Web(Lan)/RS-485/RS-232C. GPIB(not standard option).

  • Shelf・shelf brackets

    Used for the placement of specimens in the test area.
    Same as standard shelf.
    *Shelf for H-(H)30/40/V-(H)33 type is punched plate.

    Parameter
  • Cable port (without rubber plug)

    A through hole provide on the wall of chamber。
    Available dimensions: θ50/θ100
    *1.without rubber plug and with cap.
    *2.Could be added into the wall of chamber(standard position only ).Non-standard posotion please Inquire the sale company for details.
    *3.Ifseveral cable ports are installed,the sueface temperature may rise or the chamber may not be able to meet standard performance.

  • Rubber plug for cable port(200℃ only)

    Comes with the cable port. Prevents airleakage from the cable port.
    Inside diameter: θ50/θ100
    *1.The rubber plug cannot be used when operating the chamber at+200℃ or higher.

  • Automatic damper

    Automatically opens or closes synchronously with program operation for ventilation and fast cooling of chamber temperature.

  • Airflow adjuster

    Allows low air velocity in the chamber.Use when test require low airflow velocity or constant velocity.
    Setting value rang:4 levels.

  • Alarm output terminal

    Safety device.
    Purpose: This contact signal terminal is for sending error alerts to a remote location during safety actions.
    *1 Unable to be fitted together with three Color Status indicator Warning Lights .

  • Emergency push stopbutton

    Safety device,enables manual emergency stop of the chamber.
    *Purpose:Protecting the specimens and chambers.

  • Three Color Status indicator Warning Lights

    Safety device.
    Purpose: Provides a view of the chamber status from a remote location.
    With buzzer/Without buzzer
    *1:Unable to be fitted together with Alarm output terminal.
    *2:Please contact the sales company if you need to change the action and light sequence.

  • Double stacking brackets

    When stacking two chambers,this plate couples the top and bottom chambers securely.
    *Only available for GPH-(H)20 type.

  • Stand(Horizontal)

    Lift the device up,make it easier and more convenient to put in and take out specimens from the lower area inside chamber.

    Parameter
  • Stand(Vertical)

    Lift the device up,make it easier and more convenient to put in and take out specimens from the lower area inside chamber.

    Parameter
  • Stand(with door)

    Lift the device up,make it easier and more convenient to put in and take out specimens from the lower area inside chamber. More beautiful exterior with the door.
    Available for the H-2,H-3 type stand of GPH-(H)20/30.

  • Casters for stand

    Installed on stand for mobility.
    Available for stand L-2,M-2,H-2(Horizontal),and stand MV-23,MV-26(Vertical).

  • GB specifications inspection report

    Test according to GB standard and issue test report.
    *Base on the actual machine test results and for customer reference only. The report does not have the qualificationas a third-party testing agency.

  • Standard inspection report

    Test report issue by ESPEC guangdong,
    Same as standard accessories. shipped with chamber.